2、依照(zhao)被測絕(jue)緣(yuan)表(biao)的額定(ding)電(dian)(dian)壓(ya)(ya)及被測項目選定(ding)多功能(neng)試(shi)驗箱的電(dian)(dian)壓(ya)(ya)表(biao)量(liang)程,“電(dian)(dian)壓(ya)(ya)種類”開(kai)(kai)關和“測量(liang)功能(neng)”等(deng)開(kai)(kai)關放置在合(he)適(shi)位置,
例如:測500V絕緣(yuan)電(dian)(dian)阻(zu)表的開(kai)路(lu)電(dian)(dian)壓(ya)時,“量(liang)程”開(kai)關(guan)置于500V檔(dang),“電(dian)(dian)壓(ya)種類”開(kai)關(guan)置Uk檔(dang),其(qi)它開(kai)關(guan)都置于“OFF”檔(dang)。開(kai)機(ji)預熱5分鐘 。
3、接通(tong)電(dian)(dian)子(zi)式絕緣電(dian)(dian)阻表的電(dian)(dian)源(yuan)開(kai)關或啟動恒速器,儀器開(kai)始(shi)測(ce)量,待被測(ce)電(dian)(dian)壓值(zhi)顯示穩定后,就可讀取(qu)中值(zhi)、跌落(luo)、開(kai)路或峰值(zhi)電(dian)(dian)壓。
需要說明的是各開關位(wei)置放置一定要正(zheng)確,如:測量開路電壓、峰(feng)值(zhi)電壓時,
“中(zhong)值電(dian)阻”開(kai)(kai)關(guan)和“跌落電(dian)阻”開(kai)(kai)關(guan)必(bi)須放置(zhi)在“OFF”位置(zhi),測量中(zhong)值電(dian)壓時,選(xuan)定好相應的中(zhong)值電(dian)阻值,“跌落電(dian)阻”開(kai)(kai)關(guan)必(bi)須置(zhi)于(yu)“OFF”位置(zhi),
反(fan)之測量跌落電(dian)(dian)壓時,選(xuan)定好相應的跌落電(dian)(dian)阻值,而“中值電(dian)(dian)阻”開關必須置(zhi)于(yu)“OFF”位置(zhi)。
4、若要測短路電流時,需將“量程”開關置于Id 20mA或2mA檔,“測量功能”開關置于I檔,其它開關都置于“OFF”檔,即可測量短路電流。2、由于數字電壓表有較長的時間常數(約30S),故一次測量完畢后需停止恒速器轉動,或關閉電子絕緣表的電源,
待高壓跌(die)落到百(bai)伏(fu)以下(xia)時,再換量程或改變(bian)測(ce)量項(xiang)目。
3、測量過程中,如電壓表出現一明一暗閃爍時,表示輸入電壓過高或量程錯,或開關位置放置不正確,需轉換至正確量程。4、在(zai)檢(jian)(jian)定中值電阻(zu)和跌落電阻(zu)時所施(shi)加的(de)電壓,應滿足下表的(de)要求。檢(jian)(jian)測(ce)中值電阻(zu)時,“其它開關”必須置于(yu)“OFF”位置,
檢測跌(die)落(luo)電阻(zu)時,其它(ta)開關也必須(xu)置于“OFF”位置。
注:若檢某一電阻出現阻值漂移時,只要在一分鐘左右的時間不超差,仍可視為正常。電阻值 | ≤390k | ≤2MΩ | ≤10MΩ | ≤200MΩ | >200MΩ |
電壓V | 100 | 500 | 1000 | 2500 | 5000 |
5、本(ben)產品(pin)峰值電壓測量時,采用零位補(bu)償法克服峰值取樣電路中整流器的正向壓降
。故Uf檔沒有(you)輸(shu)入信號也有(you)一定大小(xiao)的(de)零位(wei)是正常現象。
直流電壓 | ≤500V: ±(0.4%讀數+0.1%滿(man)度) |
>500V: ±(0.7%讀數+0.3%滿度(du)) | |
峰值電壓 | ≤500V: ±(1%讀(du)數+0.5%滿度) |
>500V: ±(0.8%讀(du)數+0.2%滿(man)度) | |
短路電流 | ±(0.8%讀數(shu)+0.2%滿度) |
中值電阻(zu) | ±5% |
跌(die)落電(dian)阻 |